indication on dc voltmeter.
(3) Connect dc voltmeter between +5 volt test point
(2) Adjust R9801 (fig. 6) for 5.00 volt indication on
(fig. 6) and chassis ground. If dc
voltmeter does not
indicate between 4.95 and 5.05 volts, perform b(2)
(3) Adjust R9831 (fig. 6) for 15.00 volt indication on
below.
(4) Repeat technique of (2) above; dc voltmeter
indication will remain between 4.95 and 5.05 volts.
(5) Connect dc voltmeter between + 15 volt test point
14. Final Procedure. a. Deenergize and disconnect all
(fig. 6) and chassis ground. If dc
voltmeter does not
equipment and replace TI protective cover.
indicate between 14.85 and 15.15 volts, perform b(3)
b. In accordance with TM 38-750, annotate and affix
below.
DA Label 80 (US Army Calibrated Instrument). When
(6) Repeat technique of (2) above; dc voltmeter
the TI receives limited or special calibration, annotate
indication will remain between 14.85 and 15.15 bolts.
and affix DA Label 163 (US Army Limited or Special
b. Adjustments
tolerance, annotate and affix DA Form 2417 (US Army
Calibration System Rejected Instrument).