TB 11-6625-2713-50
NOTE
Interaction exists between A3A5R24 and A3A5R25 Repeat
adjustments as necessary for best compromise.
a. Performance Check
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
(1) Connect digital voltmeter (A3) to TI TP2 (fig. 8) and chassis ground, using leads
supplied with digital voltmeter. Digital voltmeter will indicate between -14.85 and -15.15
volts.
(2) Repeat technique of (1) above for TI test points and digital voltmeter indications
listed in table 7. If digital voltmeter does not indicate within limits specified, perform
appropriate adjustments listed in table 7.
18