(5) Using the same procedure, perform the
10. Final Procedure. a. Deenergize and disconnect all
test equipment, and install unit under test in protective
cover.
Table 3. Meter Accuracy Check
b. In accordance with TM 38-750, annotate and
Unit under test
Meter test set
meter
System). When unit under test cannot be adjusted
indication
Minimum
Maximum
within tolerance, annotate and affix DA Form 2417
80
78
82
(Unserviceable Test Instrument or Standard) (red tag).
60
58
62
40
38
42
20
18
22
b. Adjustments. No adjustments can be made.
6