TB 55-4920-428-35
(12) Repeat technique (6) thru (8) above, using applied pressure from dead weight tester of 500, 1500, 2500, 3500, and
4500 psi. TI high pressure gage will indicate within 50 psi of nominal pressure generated by the dead weight tester for each
indication.
(13) Disconnect dead weight tester from TI high pressure gage and reinstall in hydraulic unit.
b. Adjustments. No adjustments can be made.
13. Force Gage.
a. Performance check.
(1) Remove TI force gage from top of hydraulic top case.
(2) Attach weight holder (B11) to bottom side of TI force gage and record tare weight.
(3) Attach weights (A8) to the TI force gage weight holder to equal 10 lbs. TI force gage will indicate the reading
within 1 division plus tare weight.
(4) Attach weights to the TI force gage weight holder to equal 30, 60, and 75 lbs. TI force gage will indicate the read-
ing within 1 division for each applied weight plus tare weight.
b. Adjustment. No adjustment can be made.
Dial Indicators. See TB 9-5210-210-50.
14.
a. Performance Check. DELETE.
Change
2
27