TB 55-4920-429-35
9.
Power Supplies Voltages.
Performance Check.
a.
(A3) leads (B4) and (B5)
to the TI HI/LO test point jacks position.
Set TI 115 vac POWER switch to ON.
(2)
Set TI 28 vdc POWER switch to ON.
(3)
(4) Set TI METER SOURCE to OFF and SELF TEST Switch to
OFF position.
TI will indicate 0000 0005.
Set TI METER SOURCE to SELF TEST and SELF TEST Switch
(5)
to DISPLAY position.
TM DMM will indicate +.8.8.8.8.
Set TI SELF TEST to DM +5 vdc.
(6)
will indicate between 4.90 and 5.10 vdc.
If not, perform b(1)
and b(2) below.
TI DMM will indicate between 4.90 and 5.10 vdc.
(7)
(8) Set TI SELF TEST switch to each switch position as
listed in tables 5.0 and 6.0.
indicator will indicate within the limits specified.
13