TB 55-4920-429-35
Final procedure.
12.
De-energize and disconnect all equipment and reinstall
a.
protective cover on TI.
when all parameters are within tolerance, annotate and
b.
When the TI
affix DA Label 80 (US Army Calibrated Instrument).
receives limited or special calibration, annotate and affix DA
When the
Label 163 (US Army Limited or Special Calibration).
TI cannot be adjusted within tolerance, repair the TI in accor-
When repair is delayed for
dance with the maintenance manual.
any reason or the TI cannot be repaired with local resources,
annotate and affix DA Form 2417 (US Army Calibration System
Rejected Instrument) and inform the owner/user accordingly in
accordance with TB 750-25-1.
23