TB 9-1430-201-50
17. Final Procedure
a. Deenergize and disconnect all equipment.
b. In accordance with TM 38-750, annotate and affix DA Label 80 (US Army
Calibration System). When the TI cannot be adjusted within tolerance, annotate and affix
DA Form 2417 (Unserviceable or Limited Use tag).
SECTION VI
GENERAL RADIO, MODEL 14154A
NOTE
Unless otherwise specified, verify the results of each test and
take corrective action whenever the test requirement is not
18. Preliminary Instructions
a.
Personnel should become familiar with
sections I, II, and VI
before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2 and, for prefix B, see
19. Ratio Accuracy
a. Performance Check
(2) Position TI dials to 999(10).
(3) Adjust voltage divider (A3) to 999999X.
(4) Adjust output of voltage standard (A4) to 12 volts.
(5) Adjust lead compensator (A5) BALANCE 1 control for null on null detector (A1).
(6) Set TI and voltage divider controls to zero.
(7) Adjust lead compensator BALANCE 2 control for null on null detector.
12