TB 9-4910-555-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
TM 9-4910-571-12&P and TM 9-4910-571-34&P were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. The STE/ICE-R is reprogrammable.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc/low physical techniques.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Dc voltage1
Range: 0 to .49 V dc
Accuracy: +6% of FS
Range: .50 to 4.49 V dc
Accuracy: +3% of FS
Range: 4.50 to 45 V dc
Accuracy: +1% of FS
Dc current1
Range: 0 to 1500 A
Accuracy: (+7.0 to 8.0% of input) + (2% of FS)2
Range: 0 Ω to 4500 Ω
Accuracy: (+3.0% of indication) + (0.2% of FS)
Range: 0 Ω to 40,000 Ω
Accuracy: (+3% of indication) + ( 0.5% of FS)
Range: 40 to 500 Hz
Accuracy: (7% of input)+ (0.65% of FS)
RPM
Range: 50 to 5000 RPM
Accuracy: 0.5% of FS
See notes at end of table.