(2) Repeat technique of (1) above with + (positive) lead from digital multimeter
connected to CVA terminals N, LL, H, P, and MM. Indication on digital multimeter will
remain between +9.15 and +9.85 V.
(3) Using DATA ENTRY keys, enter 66666 on set communicator and repeat (1) and
(2) above. Digital multimeter indications will remain between -9.15 and -9.85 V.
(4) Press set communicator STOP and CLEAR keys. Enter 66664 on set
communicator and press GO key.
(5) Press set communicator STOP and CLEAR keys.
b. Adjustments. No adjustments can be made.
a. Performance Check
Figure 9. Frequency accuracy
equipment setup.