TB 9-4910-556-35
(2) Press and release STEP key on SC. SC will display:
DC CURRENT
Range: 0 to +2 A dc
Then: 0.000 A dc
(3) Set calibrator for 1.5 A dc output.
(4) Set calibrator to OPERATE mode.
(5) Set communicator will indicate between 1.46 and 1.53 A dc.
(6) Set calibrator to STANDBY mode.
b. Adjustments. No adjustments can be made.
a. Performance Check
(2) Press and release STEP key on SC. SC will display:
AC CURRENT
Range: 0 to 1.14 A rms
Then: 0.000 A ac.
(3) Set calibrator for 1 A ac, 400 Hz output.
CHANGE 1 25