(3) Adjust R41 (fig. 6) for .0 on % RPM indicator (R).
(4) Adjust R44 and R45 (fig. 6) for 100 on % RPM indicator (R).
(5) Repeat a (l) through (11) above until no further adjustments are required.
a. Performance Check
(1) Set STD DAY switch to ON and TEMPERATURE switch to OPERATE.
(3) Adjust oscillator for a 1.0 V, 70 Hz (14.286 ms) output. (Note: Set up 1.0 V
(4) Adjust AMB TEMP (oF) control to 49. If % RPM indicator does not indicate
between 100.4 and 101.3, perform b (l) and (2) below.
(5) Adjust AMB TEMP (oF) control to -41. If % RPM indicator does not indicate
between 110.7 and 111.7, perform b (3) below.
(6) Adjust AMB TEMP (oF) control to 139. If % RPM indicator does not indicate
between 92.6 and 93.6, perform b (4) below.