TB 9-4920-456-24
(10) If measuring receiver does not indicate within 0.l dB of calibration value posted
on cable, record measuring receiver indication and perform b below.
(11) On signal generator turn RF OUTPUT off.
(12) Disconnect cable from setup.
(13) Repeat steps (8) through (12) above for the second W4 and W5 cables.
(14) Ensure that signal generator RF OUTPUT is off.
(15) On measuring receiver deselect RATIO function.
(16) Connect measuring receiver sensor module directly to signal generator RF
OUTPUT.
(17) Set up signal generator frequency for 1030 MHz.
(18) On signal generator turn RF OUTPUT on.
(19) Set up measuring receiver for Tuned RF Level.
(20) On measuring receiver select RATIO function.
(21) On signal generator turn RF OUTPUT off.
(22) Connect one end of W4 cable to signal generator RF OUTPUT and connect the
other end of W4 cable to measuring receiver sensor module, using (TNC to BNC)
connector adapter.
(23) On signal generator turn RF OUTPUT on.
(24) If measuring receiver does not indicate within 0.l dB of calibration value posted
on cable, record measuring receiver indication and perform b below.
(25) On signal generator turn RF OUTPUT off.
(26) Disconnect cable from setup.
(27) Repeat (22) through (26) above for the second W4 and W5 cables.
(28) On signal generator turn RF OUTPUT off.
b. Adjustments. On TI cable, change the calibration label to read the measuring
receiver recorded values in (a) above to indicate frequency and insertion loss.
32