TB 9-4920-467-24
(6) Set the resistance standard to 900.00 Ω. If TI reading does not indicate between
899 and 901 Ω, perform b below.
(7) Press the TI ETTS BrainPak left Encoder button to return to the main POD
Alignment screen.
b.
Adjustments
(1) Select and activate Align on the UxValidator.
On-screen instructions are
displayed.
Allow the circuit to stabilize for
2 minutes.
(3) Select and activate OK. The TI will measure the resistance.
When the
measurement is completed, a second on-screen instruction is displayed.
Set the resistance standard to 800.00 Ω. Select and activate OK.
(4)
(5) When both resistance measurements are complete, the alignment progress bar
will show 100%. The TI will calculate the alignment constants. The TI will enter a live-
reading performance check mode.
(6) Press the TI ETTS BrainPak left Encoder knob to save the alignment
constants.
(7)
a.
Performance Check
(1) Select and activate Jrc from the Circuit menu on the UxValidator.
Jrc will
appear in the Circuit field.