TB 9-4920-467-24
(3) The preferred voltage output of +17.336 V dc for Vt_DC is shown in the output
field on the UxValidator. Using the + and - buttons on the screen, adjust the TI's output so
that the voltage displayed on the multimeter(s) matches the on-screen voltage. The
resolution of the adjustment can be changed by selecting Fine, Medium, or Coarse.
(4) When the voltage displayed on the multimeter matches the voltage displayed in
the output field on the UxValidator, Vt_DC is properly adjusted.
(5) Select and activate Done (in the upper right corner of the screen) on the
UxValidator.
(6)
a.
Performance Check
(1) Select and activate Vc_DC from the Circuit menu on the UxValidator. Vc_DC
will appear in the Circuit field.
(2)
(3)
(4)
Select and activate Test from the POD Alignment screen on the UxValidator.