TB 9-4931-217-40
MAIN DIAL to 10.0000.
(c)
(d) D-Q multiplier switch to .001X.
(2) Connect capacitance standard to TI UNKNOWN terminals 1 and 2.
(3) Adjust D-Q dial for best null indication on ac generator detector meter.
(4) Record D-Q dial indication and subtract from all subsequent measurements.
connect to TI UNKNOWN terminals 1 and 2.
(6) Adjust resistance standard controls to 15.0 ohms.
(7) Adjust ac generator detector DET GAIN control for a convenient indication on
null detector.
NOTE
A broad undefined change corresponding to MAIN DIAL
change in normal.
(8) Alternately adjust MAIN DIAL and D-Q dial for a null indication on ac
generator detector null meter.
(9) Repeat (7) and (8) above for best null indication on null meter. D-Q dial will
indicate within limits specified in table 10.
(10) Rotate ac generator detector DET GAIN control fully ccw
(11) Repeat technique of (6) through (10) above, using settings and indications listed
Test instrument D-Q dial indications
Test instrument
Resistance standard
D-Q switch
indications
settings
(ohms)
Min
Max
.001X
15.0
9
11
.001X
31.8
19
21
.001X
47.7
29
31
.001X
63.7
39
41
.001X
79.6
49
51
.001X
96.5
59
61
.001X
111.4
69
71
.001X
127.3
79
81
.001X
143.2
89
91
.001X
159.2
99
101
.01X
159
9
11
.01X
318
19
21
.01X
477
29
31
.01X
637
39
41
.01X
796
49
51
.01X
955
59
61
22