TB 9-4931-383-50
Figure 1. Measurement of 1.018 + ∆E output of dc transfer standard - equipment setup.
NOTE
When adjusting for a null indication on null detector, always
start on highest range of null detector and proceed down range
to required sensitivity. When reading is complete, return null
detector to high range.
(2) Adjust dc transfer standard (A2) for 1.1018 + ∆E mean output and set ∆E dials
to last three digits of standard cell (A6) certificate value.
6