TB 9-4931-488-24
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Performance specifications 1
Test instrument parameters
Internal reference
Range: 5 MHz
Accuracy: + 1 ppm
Drift:
< 0.05 ppm per hr
Range: 0.5 to 512 MHz
0.5 to 1024 MHz (option 002)
Accuracy: Counter resolution ( 1 count)
+ reference error 2 ppm (locked mode)
Stability: < 10 ppm/10 minute (unlocked mode)
< 0.1 ppm for + 5% to 10% line voltage
change (locked mode)
≥30 dB below fundamental, 0.5 to 512 MHz
Harmonics (at 1 V,+ 10 dB output range and below)
≥12 dB below fundamental, 512 to 1024 MHz
(option 002)
≥20 dB below fundamental, 512 to 1024 MHz
Output Level (Top 10 dB of Vernier Range) 2
Options
Output Level (dBm)
Accuracy (dB)
(MHz)
Basic, 001, and H66
0.5
to
512
+19
to
-7
1.5
-7
to
-47
2.0
3
- 47
to
-137
2.5
002
0.5
to
64
+18.5 to
-7
1.5
-7
to
-47
2.0
3
- 47
to
-137
2.5
64
to
512
+18.5 to
-7
2.0
-7
to
-47
2.5
3
-47
to
-137
3.0
512 to
1024
+13
to
-7
3.0
-7
to
-47
3.5
3
-47
to
-127
4.0
+1.75/-2.25
003
0.5
to
512
+18.5 to
-7
+2.25/-2.75
7
to
-47
3
+2.75/-3.25
47
to
-137
002/003
0.5
to
512
+18
to
-7
+2.0/-3.0
-7
to
-47
+2.5/-3.5
-47
to
-1373
+3.0/-4.0
512
to
1024
+12
to
-7
3.5
-7
to
-47
4.0
-47
to
-1283
4.5
004
0.5
to
512
+15
to
-10
1.5
-10
to
-50
2.0
-50
to
-142 3
2.5