TB 9-4931-488-35
Test instrument
Spectrum analyzer
Max Mkr1
Stop
Start
Range MHz switch
Frequency MHz
indication
Freq
Freq
setting
display
(Hz)
(Hz)
(MHz)
below fundamental
1 - 0.5
.5
400
K
2.6
M
-30
2-1
1
900
K
5.1
M
-30
4-2
2
1.9 M
11
M
-30
8-4
4
3.9 M
21
M
-30
16 - 8
8
7.9 M
41
M
-30
32 - 16
16
15
M
81
M
-30
64 - 32
32
31
M
161
M
-30
128 - 64
64
63
M
321
M
-30
256 - 128
128
127
M
640
M
-30
512 - 256
256
250
M
1.3
G
-30
512 - 256
512
500
M
2.6
G
-30
NOTE
For TI's containing option 002, perform (9) through (13) below.
For all others set RF OFF/ON switch to OFF and proceed to
(9) Press spectrum analyzer Marker, [Off].
(10) Position controls as listed in (a) through (d) below.
(a)
RANGE MHz switch to 1024-512 MHz.
(b)
OUTPUT LEVEL switch to +10 dBm.
(c)
RF OFF/ON switch to ON.
(d)
COUNTER MODE EXPAND X10 pushbutton released.
(11) Adjust FREQUENCY TUNE control for a frequency MHz display indication of
1024 MHz and OUTPUT LEVEL vernier control(s) fully cw.
(12) Set spectrum analyzer to TI frequency, set power reference Marker, Peak
Search,
Marker,
[Delta], then tune to
harmonic frequency listed in
table 4. Harmonic
frequency Mkr1 indication will be within limits listed in table 4.
(13) Repeat technique of (9) through (12) above as necessary for remaining rows in