Table 3. Power Output (Model 86222A)
Frequency Range (Model 86290A)

Calibration Procedure For Sweep Oscillator Hewlett-Packard, Models 8620A And 86920C With Rf Plug-Ins, Models 86200 And 86300 Series; And Modules 8621A And 8621B Page Navigation
  11    12    13    14    15  16  17    18    19    20    21  
img
. . .
TB 9-4931-495-24
20. Frequency Range (Model 86222A)
a. Performance Check
(1) Connect RF OUTPUT to microwave frequency counter, using attenuator (fixed).
(2) Position controls as listed in (a) through (f) below:
(a) Set POWER LEVEL control to +13 dBm.
(b) BAND switch to .01-2.4 GHz.
(c) START MARKER pointer to 20 MHz.
(d) CW MARKER pointer to 1.2 GHz.
(e) STOP MARKER pointer to 2.4 GHz.
MODE switch to MANUAL.
(f)
Figure 4. Model 86222A - top view.
(3) Press CW pushbutton.
(4) Set CW MARKER pointer to 10 MHz. If microwave frequency counter does not
indicate between 0.00 and 20 MHz, perform b below.
(5) Repeat (4) above at 1.2 and 2.4 GHz. If microwave frequency counter does not
indicate CW MARKER pointer setting ±10 MHz, perform b below.