TB 9-4931-504-50
Test instrument
Performance specifications
HORIZONTAL SYSTEM
Delaying time base:
100 s/div to 10 ns/div
Time base range
Time base accuracy
Within 2%, excluding first 1/2 div of displayed sweep
Delay zero range
0-9 div or more
Delay time multiplier
Each turn equal to 1 crt division
Delay accuracy
Within 1% of full screen when measured between 1st and 10th crt
division
Delayed time base:
100 s/div to 100 ps/div in 19 steps of 1, 2, 5 sequence
Range
Accuracy
3%, excluding first 1/2 div of displayed sweep
Variable
At least 2.5:1
Time base display modes:
Conventional display, max lead time, left intensified dot indicates
time zero (multiplier zero), right intensified dot indicates point at
which delayed sweep starts. Time between dots is read from crt or
delay time multiplier dial.
Delayed sweep starts immediately at end of delay time. Set by
Delayed time base
delay plus delay time multiplier.
Triggering:
Amplitude range
External
10 mV to 2 V p-p
Internal
50 mV to 2 V p-p
51Ω +10%, ac coupled
Normal triggering modes:
Sine waves
150 kHz to 100 MHz
Pulse
10 Hz to 100 MHz
10 mV/s
Minimum rise rate
AUTO TRIG modes:
105 kHz to 100 MHz
Sine waves
10 mV p-p at 100 MHz (ext)
Minimum amplitude
1 kHz to 100 MHz
Pulse
10 ns at 1 kHz
10 mV/s
Minimum pulse width
Minimum rise rage
100 MHz to 1 GHz
HF SYNC mode:
Sine waves
Scan controls:
24-40 Hz repetition rate
Repetitive
1 sweep/single sweep start pushbutton pressed
Single sweep
slightly greater than 10 div
1 V/div 5%
Manual scan range
Ext scan
150 V
Max sensitivity
Max input voltage
1 V/div 5%
Horizontal output signal
Amplitude