TB 9-4931-537-35
(2) Press multimeter FUNCTION/RANGE DCI
pushbutton and set Guard
switch to To LO (depressed position).
(3) Preset (in standby mode) calibrator for a 100 A output. Press Currnt
Output softkey until "Currnt Output NORMAL" is displayed.
(4) Set calibrator to Operate. Record multimeter indications in space provided.
(5) Set calibrator to remaining (m) A dc outputs listed in table 1.
Record
multimeter indications in space provided.
(6) Set calibrator to STANDBY, then press RESET pushbutton.
g. Ac Current. Calibrator, John Fluke, Model 5700A/( ), Multimeter, Hewlett
Packard, Model 3458A and Amplifier, John Fluke, Model 5725A/( ).
(1) Press multimeter FUNCTION/RANGE ACI pushbutton.
(2) Set calibrator to (, m) A, (k)Hz outputs listed in table 1. Record multimeter
indications in space provided.
(3) Set calibrator to STANDBY, then press RESET pushbutton.
h. Resistance :
Calibrator, John Fluke, Model
5700A/( ), and Hewlett
Packard,
Model 3458A.
(1) Connect equipment as shown in figure 3. Set multimeter Guard switch to
Open.
FUNCTION/RANGE blue colored pushbutton then the OHM/OHMF pushbutton. Set
O Comp to ON by pressing MENU Offset Comp Ω pushbutton, followed by
NUMERIC/USER 1 and Enter pushbuttons.
8