TB 9-4931-540-24
(g) SWEEP ENTRY MANUAL to 60 Hz.
(h) ENTRY CF STEP SIZE to 60 Hz.
(i) TRACE CLEAR A.
(3) Average TI marker amplitude indication will be <-120 dBm.
(4) Press SWEEP ENTRY MANUAL key and then ENTRY STEP
key. Average
TI marker amplitude indication will be <-120 dBm.
(5) Press ENTRY STEP
key. Average TI marker amplitude indication will
be <-120 dBm.
(6) Press ENTRY STEP
key. Average TI marker amplitude indication will
be <-120 dBm.
(7) Repeat (6) above six times.
(8) Press keys and enter values using ENTRY keys as listed in (a) through (c) below:
(a) SWEEP ENTRY MANUAL to 5 kHz.
(b) ENTRY CF STEP SIZE to 5 kHz.
(c) TRACE CLEAR A.
(9) Repeat (3) through (7) above.
(10) Press keys and enter values using ENTRY keys as listed in (a) through (c) below:
(a) SWEEP ENTRY MANUAL to 100 kHz.
(b) ENTRY CF STEP SIZE to 100 kHz.
(c) TRACE CLEAR A.
(11) Repeat (3) through (7) above.
(12) Press keys and enter values using ENTRY keys as listed in (a) through (c) below:
(a) SWEEP ENTRY MANUAL to 1 MHz.
(b) ENTRY CF STEP SIZE to 1 MHz.
(c) TRACE CLEAR A.
(13) Repeat (3) through (7) above.
(14) Press keys and enter values using ENTRY keys as listed in (a) through (c) below:
(a) SWEEP ENTRY MANUAL to 10 MHz.
(b) ENTRY CF STEP SIZE to 10 MHz.
(c) TRACE CLEAR A.
(15) Repeat (3) through (6) above.
16