TB 9-4931-541-40
Table 3. Calibration Data Chart for Dial Indication Calibrator (DIC)
TI contact surface direction
Electronic Linear Transducer Display Unit indication
TI
display
1st Pass
2nd Pass
Minimum
Maximum
UP
DOWN
0.050000
0.049900
0.050100
UP
DOWN
0.100000
0.099900
0.100100
UP
DOWN
0.200000
0.199900
0.200100
UP
DOWN
0.300000
0.299900
0.300100
UP
DOWN
0.400000
0.399900
0.400100
UP
DOWN
0.500000
0.499900
0.500100
UP
DOWN
0.600000
0.599900
0.600100
UP
DOWN
0.700000
0.699900
0.700100
UP
DOWN
0.800000
0.799900
0.800100
UP
DOWN
0.900000
0.899900
0.900100
UP
DOWN
1.000000
0.999900
1.000100
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.