TB 9-4935-364-24
Test instrument
TEST VOLTAGE
switch settings
Min
Max
10 V
9.8
10.2
11 V
10.78
11.22
12 V
11.76
12.24
13 V
12.74
13.26
14 V
13.72
14.28
15 V
14.70
15.30
16 V
15.68
16.32
17 V
16.66
17.34
18 V
17.64
18.36
19 V
18.62
19.38
20 V
19.6
20.4
30 V
29.4
30.6
40 V
39.2
40.8
50 V
49.0
51.0
60 V
58.8
61.2
70 V
68.6
71.4
80 V
78.4
81.6
90 V
88.2
91.8
100 V
98.0
102.0
100 V
98.2
102.0
110 V
107.8
112.2
120 V
117.6
122.4
150 V
147.0
153.0
300 V
294.0
306.0
400 V
392.0
408.0
500 V
490.0
510.0
600 V
588.0
612.0
700 V
686.0
714.0
800 V
784.0
816.0
900 V
882.0
918.0
1000 V
980.0
1020.0
(7) Set function switch to DISCHARGE.
b. Adjustment. Adjust R140 (fig. 1) until multimeter indicates 200 V dc (R).
9. Meter Tracking and Resistance Accuracy
a. Performance Check
(1) Connect resistance standard No. 1 HIGH and LO terminals to TI + (positive) and
- (negative) UNKNOWN terminals respectively and adjust resistance standard for 500 k .
(2) Set TEST VOLTAGE switches to 1-0-0V.
(3) Set function switch to MEASURE.
(4) Adjust resistance standard for a TI meter reading of 0.5. If resistance standard
No. 1 does not indicate between 485 and 515 k , perform b below.