TB 9-4935-364-24
(5) Adjust resistance standard for a meter reading of 1. Resistance standard will
indicate between 960 k and 1.04 M .
(6) Set function switch to DISCHARGE.
(7) Connect resistance standards No. 1 and No. 2 in series to TI + (positive)
and - (negative) UNKNOWN terminals. Adjust resistance standards for a resistance of 5 M .
(8) Connect GUARD terminal in series with resistance standards GUARD terminals.
(9) Set function switch to MEASURE.
(10) Adjust resistance standards for a TI meter reading of 5. If resistance standards
do not indicate between 4.4 and 5.6 M , perform b below.
(11) Set function switch to DISCHARGE.
(12) Adjust resistance standard for setting of 50 k .
(13) Adjust TEST VOLTAGE switches for 1-0-V.
(14) Set function switch MEASURE.
(15) Adjust resistance standard for 0.5 indication on TI meter. Resistance standards
will indicate between 48.5 and 51.5 k .
WARNING
Set function switch to DISCHARGE, before disconnecting or
connecting resistance standards to TI in table 5 below.
(16) Repeat technique of (11) through (15) above for TI settings listed in table 4.
Resistance standards indications will be within limits specified.
Resistance standards1
Test instrument
TEST
Multiplier
VOLTAGE
Meter
Preliminary
Final resistance setting
switch
( )
switch
indications
settings
settings
( )
( )
settings
Min
Max
100 k
1-0-V
0.7
70 k
67.62 k
72.38
k
100 k
5-0-V
1.0
100 k
96
k
104
k
100 k
5-0-V
2.0
200 k
188
k
212
k
100 k
5-0-V
5.0
500 k
440
k
560
k
1 M2
1-0-0V
1.0
1 M
0.96 M
1.04
M
10 M2
1-0-0V
1.0
10 M
9.6 M
10.4
M
100 M2
1-0-0V
1.0
100 M
96
M
104
M
1 G2,3
1-0-0V
Min.
Max.
1 G
---
---
0.96
1.04
10 G2,3
1-0-0V
Min.
Max.
10 G
---
---
0.96
1.04
1Connect
resistance standards (No. 1 through No. 6) as required.
2If
an out-of-tolerance condition exists, the SET
controls may require adjustment. Set function switch to DISCHARGE.
Remove connections to TI UNKNOWN INPUT. Set function switch to MEASURE. Adjust controls for indication. Set
function switch to DISCHARGE. Reconnect resistance standards to original setup. Set function switch to MEASURE.
3Connect resistance standard No. 5 to UNKNOWN terminals with resistance standard middle terminal connected directly to
TI ground terminal.
7