TB 9-4935-552-50-2
9. Intensity
NOTE
Do not proceed until the TI and the radiometer A2 have
warmed up at least 30 minutes.
a. Recenter the filament image, if necessary, and set cover back on TI. Do not tighten
cover screws.
b. Connect passive Probe A to self-test reference 1A7 (7 VAC left/right).
Program
SMPV12@K.
c. Test results display (TRD) shall indicate 19.80 65 VPP.
d. If TRD is less than 19.15 or greater than 20.45 VPP, record the percent correction
and apply to the TRD VPP indications produced hereafter.
e. Connect radiometer A2 OUT to
waveform converter using passive probe A and
Figure 5. AC/DC and passive probe A.
10