TB 9-5120-202-35
Table 3. Minimum Specifications of Equipment Required - Continued
Minimum use
Manufacturer and model
Common name
specifications
(part number)
TORQUE CELL NO. 3
Range: 0 to 100 ft-lb
Lebow Associates, Model 2133-125
(MIS-26485, Type 1, CL3)
from 20% FS to FS
0.5% of 20% FS below
20% FS
TORQUE CELL NO. 4
Range: 0 to 500 ft-lb
Lebow Associates, Model 2133-126
(MIS-26485, Type 1, CL4)
from 20% FS to FS
0.5% of 20% FS below
20% FS
TORQUE CELL NO. 5
Range: 0 to 1000 ft-lb
Lebow Associates, Model 2133-127
(MIS-26485, Type 1, CL5)
from 20% FS to FS
0.5% of 20% FS below
20% FS
TORQUE TESTER
Range: 20 to 200 in-oz.
AWS3002 (MIS-45850)
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in the manufacturer's manual for these TIs.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.