Figure 2. Test Instrument Orientation (0 to 45°).
Superset Alignment

Calibration Procedure For Digital Protractors, Kell-Strom, Models Pro360 And Pro3600 Page Navigation
  1    2    3    4  5  6    7  
img
. . . . . . .
TB 9-5210-215-24
9. Angle Indication Greater Than 45°
a. Performance Check
(1) Setup equipment on level surface plate as shown in figure 3 using the 5-inch
sine plate.
Figure 3. Angle Indication (> 45°) Equipment Setup.
(2) Ensure TI HOLD and ALT ZERO functions are deactivated.
(3) Check TI in positions 5 through 8 (fig. 4). TI will indicate within limits specified
in table 4. If not, perform b below.
Figure 4. Test Instrument Orientation (>45°).