TB 9-5210-219-40
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
were used as the prime data sources in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None
b. Time and Technique. The time required for this calibration is approximately
1 hour, using the physical technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameter
Performance specifications
o
o
Angle indication
Range: -45 to 45
Accuracy: -10o to 10o, 0.1o; -10o to 45o, 1.5% rdg
(rounded to next greater 0.1o )
o
o
Range: -60 to 60
Accuracy: -50o to 50o, 0.1o; -50o to -60o, 50o to 60o, 0.3o
o
o
Over range
Range: -45 to 45
indicates over range between -50o to -60o, 50o to 60o
Accuracy:
Range: -60o to 60o
Accuracy: indicates over range <-60o, >60o
o
o
Alternate reference
Range: -45 to 45
Accuracy: -10o to 10o, 0.1o; -10o to 45o, 1.5% rdg
(rounded to next greater 0.1)
o
o
Range: -60 to 60
Accuracy: -50o to 50o, 0.1o; -50o to -60o, 50o to 60o, 0.3o
o
o
o
o
Independent sensor
Range: -45 to 45 ; -60 to 60