TB 9-5855-1890-35
5. Accessories Required. Common usage accessories not listed in this calibration
procedure and peculiar accessories listed in table 3 are issued as indicated in paragraph
4 above.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 0 to 5 C differential
DIFFERENTIAL
Omega Engineering 0.020
Accuracy: 0.11 C
THERMOCOUPLE SET
Type K (7917040)1
Range: 0 to 200 V
DIFFERENTIAL
John
Fluke,
Model
887AB/AN
Accuracy: 2 V
VOLTMETER
(887AB/AN)
DIGITAL VOLTMETER
Range: 0 to 6 V dc
Hewlett-Packard, Model 3490AOPT060
Accuracy: 0.2%
(3490AOPT060)
1Part
of calibration kit 7917034, limited deployed.
Table 3. Accessories Required
Description
Common name
(part number)
ADAPTER1
Offset voltage pin (7917036)2
COTTON TIP APPLICATOR  3
Swap with cotton tip on one end
HOLDING CLAMP1
Mounting post clamp (7917039) 2
LABORATORY CLAMP1
Laboratory clamp (MS-36012-1) 2
LEAD
42-in., single banana to test hook (red) (7917035-2) 2
LEAD
42-in., single banana to test hook (black) (7917035-1) 2
PASTE REMOVAL SOLVENT3
WD40 or equivalent
THERMOCOUPLE MOUNTING POST1
Mounting post with base (7917037) 2
THERMALLY CONDUCTIVE PASTE3
Omegatherm 201 or equivalent
1Two
required.
2Component
part of calibration kit 7917034, limited deployed.
3Additional equipment required.
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by
common name as listed in tables 2 and 3.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 9-5855-255-14 for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.