• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Figure 2. Equipment setup. - TB-9-5855-1890-350007
Calibration

TB-9-5855-1890-35 Thermal Collimator SM-D-805691(P/O AN/TAM-3 and AN/TAM-3A) and Infrared Collimator SM-D-969722 Manual
Page Navigation
  1    2    3    4  5  6    7    8    9  
TB 9-5855-1890-35
Figure 3. Temperature controller - heater plate.
l. Install shroud on TI.
NOTE
Verify that differential voltmeter operates properly in
battery operate mode.
m. Pull leads through exit port and connect to differential voltmeter observing
indicated polarity.
n. Open shroud access door (top) and remove 4 screws from top of cover plate of TI
temperature controller and remove cover plate.
o. Place offset voltage pin adapters on pin J7 (black) and J9 (red) (fig. 1). Connect
leads to proper pin adapter (black to black, red to red).
p. Draw leads through side access port of shroud and connect to digital voltmeter.
q. Connect temperature controller to 115 V ac power source.
NOTE
Ensure that power switch is in OFF position.
6