TB 9-5855-1892-24
(5) Place tip of thermocouple No. 1 (TC1) in thermal conductive paste against
bottom of HEATER PLATE (INSIDE) BAR TARGET AMBIENT PLATE (fig. 6).
CAUTION
Avoid scratching internal edges of BAR TARGET (fig. 6)
during application and removal of thermal conductive paste.
c. Perform steps (1) through (4) below:
(2) Increase tension on thermocouple No. 1 (TC1) to produce a moderate bow in
thermocouple wire by sliding laboratory clamp (fig. 6) forward (toward temperature
controller) in mounting post clamps; then tighten mounting post clamps (fig. 5).
(3) If necessary, physically position thermocouple TC1 on HEATER PLATE to the
(4) Daub a small amount of thermally conductive paste on exterior of TI BAR
TARGET adjacent to placement of TC1 and place tip of thermocouple No. 2 (TC2) in the
7