TB 9-5855-1892-35
a. Set target select knob to BAR.
b.
Connect equipment as
shown in figure 5 and as
described in
(1) through (5) below:
DIFFERENTIAL
DIFFERENTIAL
THERMOCOUPLE
THERMOCOUPLE
TC1
LEADS
DIFFERENTIAL
LABORATORY
THERMOCOUPLE
MOUNTING
MOUNTING
CLAMP
TC2
POST CLAMP
POST
(1) Install thermocouple MOUNTING POST (pair) (fig. 5) on base of TI
(approximately center of second slot pair from exit port).
(2) Install thermocouple harness wire to thermocouples (connectors are keyed and
labeled).
(3) Clamp DIFFERENTIAL THERMOCOUPLE set (pair) in LABORATORY
CLAMPS and install on thermocouple MOUNTING POSTS, using MOUNTING POST
CAUTION
Avoid scratching internal edges of BAR TARGET (fig. 6) during
application and removal of thermal conductive paste.