TB 9-5895-1913-40
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
AC MEASUREMENT STD
Frequency Range:
Fluke Model 5790A
(MEASUREMENT STANDARD)
10 Hz to 100 kHz
(13534003)
Range: 2.24 Vrms (+20 dBm)
to 0.999 mVrms (-47 dBm).
Accuracy: 0.144%
(on 2.2 mV range only 0.172%
at 1 kHz, 0.281% at 20 kHz,
0.370% at 100 kHz)
FREQUENCY COUNTER
Frequency Range:
Fluke, Model PM6681/656
0 kHz to 250 MHz
(PM6681/656)
TIME FREQUENCY
WORKSTATION
MEASURING RECEIVER /
Spectral purity measurement Measuring receiver system N5530S
SENSOR MODULE
Range:
consisting of: Spectrum Analyzer,
Carrier from 10 Hz to 4 GHz
Agilent Model E4440A (E4440A),
Amplitude:
and Sensor Module, Agilent Model
+24 dBm to -75 dBm
N5532A opt. 504 (504)
Frequency measurement:
Range 1 GHz to 4 GHz
Accuracy: See TIME
FREQUENCY
WORKSTATION
AM rate measurement:
Carrier: 30 MHz
Range: 1 kHz to 220 kHz
Accuracy: 0.001136%
AM depth measurement:
Carrier: 125 MHz to 1 GHz
Range: 80%
Accuracy: 0.78125%
FM rate measurement:
Carrier: 125 MHz to 1 GHz
Range: 1 kHz to 200 kHz
Accuracy: 0.00083%
FM deviation: measurement:
Carrier: 125 MHz to 1 GHz
Range: 100 kHz to 1 MHz
Accuracy: 0.75% (1%)
Tuned RF Level
Measurement :
Range: -37 to -95 dBm
Accuracy:
-57 dBm 0.2 dB
-66 dBm 0.2 dB
-75 dBm 0.5 dB
-85 dBm 0.5 dB
-95 dBm 1.5 dB
4