TB 9-5975-219-24
d. Unless otherwise specified, all controls and control settings refer to the TI.
e. Unless otherwise specified, on screen controls will be accessed by using the trackball
controls to move the arrow over the icon, virtual instrument controls, etc., and clicking the
left button.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
NOTE
Field-Calibration in Section IV should be performed if TI fails
any parameter.
a. Connect TI connectors to ATE connectors as listed in (1) through (5) below:
(1) TI J1 to ATE J1 DCA.
(2) TI J2 to ATE J2 TK.
(3) TI J3 to ATE J3.
(4) TI J5 and TI J6 to ATE J6 ICE Box COMM3.
(5) TI J4 to ATE J4 Probe A and B using probe cable (13580845). (Red probe to A
and black probe to B.)
b. Connect ATE INSTRUMENT CONTROLLER COMM1 connector to Instrument
controller COM1 serial port.
c. Ensure that the PROGRAMMING ON/OFF switch located on the side of the ATE is
set to OFF.
d. Start Hyper Terminal with settings 19200 bits per second, 8 data bits, none parity,
1 stop bit, and none flow control as shown in (1) through (6) below.
(1) Open Windows Explorer.
(2) Search for hypertrm.exe.
(3) Double click on hypertrm.exe.
(4) On the Connection Description window type MSDICE in the Name: field and
click on the OK button.
(5) On the Connect To window, ensure that COM1 is selected in the Connect
using: field, and click on the OK button.
(6) On the COM1 Properties window make the following entries in the appropriate
fields as listed in (a) through (f) below:
(a)
Bits per second: 19200.
(b)
Data bits: 8.
(c)
Parity: None.
(d)
Stop Bits: 1.