TB 9-5975-219-24
(8) Set calibrator to standby.
b. Adjustments.
If TI fails any parameter, perform Field-Calibration in
a. Performance Check
(1) Select the HyperTerminal window, and at the ? prompt, enter S32 on the
instrument controller keyboard (do not press enter).
(2) Set calibrator output to 60 Hz, the first values listed in table 4. If TI indications
are not within the limits in table 4, perform b below.
(3) Set the calibrator output to standby.
(4) Select the HyperTerminal window and press the instrument controller keyboard
Enter key.
(5) Repeat technique of (1) through (4) above for remaining TI hyper-terminal range
setting commands and calibrator output settings in table 4.
T
TI hyper-
Hyper-terminal
o
Calibrator
d Test
terminal
est instrument indications
utput
(V)
escription
range
settings
(pin numbers)
setting
(V ac)
Min
Max
command
TK AC Volts (J4A_B)
S32
21
19.74
22.26
Vehicle Specific ACV (L_W)
S33
21
19.74
22.26
TK AC Current (J2A_B)
S34
2.1
1.974
2.226
TK AC Current (J3A_B)
S35
2.1
1.974
2.226
Alt. Output Current Sense (n_p)
S36
2.1
1.974
2.226
(6) Set calibrator to standby, and disconnect equipment setup.
b. Adjustments. If TI fails any parameter, perform Field-Calibration in Section IV.
a. Performance Check
(1) Select the HyperTerminal window, and at the ? prompt, enter S18 on the
instrument controller keyboard (do not press enter).
(2) Connect resistance standard HIGH and LOW connectors to the ATE SINGLE
POINT + and - banana connectors located on the side of the ATE.
(3) Set resistance standard to the first values listed in table 5. If TI indications are
not within the limits in table 5, perform b below.
(4) Repeat technique of (1) through (3) above for remaining settings in table 5.