TB 9-5985-314-35
(45) Connect TI into figure 6 equipment setup as listed in (a) through (c) below:
(a) INPUT PORT 1 to POINT A.
(b) OUTPUT PORT 3 to POINT B.
(c) OUTPUT PORT 2 to 50 Ω termination.
(46) Record receiver system indication in OUTPUT PORT 3 column of table 9.
remaining settings listed in table 9. Measure and record insertion loss for OUTPUT
(48) Algebraically calculate the difference between the values recorded in table 9 for
OUTPUT PORT 2 and OUTPUT PORT 3. Record difference in OUTPUT PORT
TRACKING column of
table 9. Output port
tracking will be within the
limits specified in
b. Adjustments.
Prepare charts (if required) similar to tables 7, 8, and 9 for TI
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.