TB 9-5985-314-35
Connect TI into figure 7 equipment setup as listed in (a) through (c) below:
(a) INPUT PORT 1 to POINT A.
(b) OUTPUT PORT 2 to 50Ω termination.
(c) OUTPUT PORT 3 to POINT B.
(33) Record measuring receiver indication in insertion loss for OUTPUT PORT 3
column in table H2. Insertion loss will be equal to or less than the limits listed in
(34) Repeat steps (20) through (24) and (26) through (30) for remaining frequencies
in Table H1. Record measuring receiver indication in insertion loss for OUTPUT PORT 3
column in table H2. Insertion loss will be equal to or less than the limits listed in
(35) Algebraically calculate the difference between the values recorded for OUTPUT
PORT 2 and
OUTPUT PORT 3 in
table H2. Record results in
OUTPUT PORT
TRACKING column in
table H2. Output port
tracking will be within the
limits specified in
b. Adjustment. Prepare a
chart (if required) similar to
table H2 for TI
frequency range.
2. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
H-6