TB 9-5985-316-24
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Set TI DB dials to 0.
8. Attenuator Range and Accuracy
a. Performance Check
.
(1) Set decade resistor to 600
(2) Energize equipment and allow sufficient time for warm-up.
(3) Adjust calibrator for an indication of 25 V dc on multimeter.