Figure 1. Attenuator range and accuracy.
Figure 2. Frequency response.

Calibration Procedure For Variable Attenuator Part Number C-8756 Page Navigation
  1    2    3  4  5    6  
img
. . . . .
TB 9-5985-316-24
NOTE
Power capacity of calibrator is 1.5 W maximum. Do not apply
more than 1.5 W input which corresponds to approximately
30 V dc or rms. Damage may also occur to TI by applying input
power to the output terminal.
(4) Set TI 0 to 10 DB dial to 1. Multimeter will indicate between 22.148 and 22.415
V dc.
(5) Set TI DB dials to positions listed in table 3. Multimeter will indicate within
limits specified.
Table 3. Attenuator Accuracy
Test instrument DB dial settings
Multimeter indications (dc)
0 to 10 dial
10 to 100 dial
Min
Max
2
0
19.739
19.977
3
0
17.592
17.805
4
0
15.679
15.869
5
0
13.974
14.143
6
0
12.455
12.605
7
0
11.100
11.234
8
0
9.8930
10.012
9
0
8.8171
8.9236
10
0
7.8583
7.9531
0
10
7.8583
7.9531
0
20
2.2750
2.7250
0
30
0.71942
0.86172
0
40
0.22750
0.27250
0
50
0.07194
0.08617
0
60
0.02275
0.02725
0
70
0.007194
0.008617
0
80
0.002275
0.002725
0
90
0.000719
0.000862
0
100
0.000228
0.000273
b. Adjustments. No adjustments can be made.
9. Frequency Response
a. Performance Check
(1) Connect equipment as shown in figure 2.