TB 9-6625-011-24
(5) Adjust R7 (R626 for TIs with amplifier board A6, P/N 03400-66512) (fig. 1) until
TI indicates 0.01 V (1 /10 scale) (R).
(6) Set RANGE switch to .01 VOLTS.
(7) Repeat (3) through (6) above until no further adjustment is necessary.
(8) Set calibrator for a 1.0 V, 400 Hz output. Adjust R104 (fig. 1) until TI indicates 1 (R).
(9) Set TI RANGE switch to .001 VOLTS. Set calibrator wideband output to 400 Hz
and adjust amplitude for a full scale reference indication. Press NEW REF on calibrator.
(10) Set calibrator frequency to 10 MHz. Adjust C405 (fig. 1) for the 400 Hz full scale
reference established in b (9) above (R).
(11) Vary calibrator frequency between 3 and 10 MHz and if required, readjust C405
(fig. 1) for best in-tolerance condition.
(12) Set TI RANGE switch to 1 VOLTS. Set calibrator wideband output to 400 Hz
and adjust amplitude for a full scale reference indication. Press NEW REF on calibrator.
(13) Set calibrator frequency to 100 kHz. Adjust C102 (fig. 1) for the 400 Hz full
scale reference established in b (12) above (R).
7