TB 9-6625-011-35
(12) Set calibrator for an initial 0.9 V, 1 kHz wideband output, then adjust for the TI
indication recorded for table 3 (0.9 V).
(13) Press calibrator NEW REF pushbutton, then set frequency to 100 kHz. Adjust
C102 (fig. 1) for the 1 kHz reference established in b (12) above (R).
(14) Vary calibrator frequency between 100 kHz and 10 MHz and, if required,
readjust C102 (fig. 1) for best in-tolerance condition.
(15) Set calibrator for an initial 0.3 V, 1 kHz wideband output, then adjust for the TI
indication recorded for table 3 (0.3 V).
(16) Press calibrator NEW REF pushbutton, then set frequency to 3.0 MHz. Adjust
C303 (fig. 1) for the 1 kHz reference established in b (l5) above (R).
(17) Vary calibrator frequency between 3 and 10 MHz and, if required, readjust
C303 (fig. 1) for best in-tolerance condition.
Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA Label/Form in accordance with TB 750-25.