TB 9-6625-096-50
SECTION IV
NOTE
Unless otherwise specified, verify the results of each test and
take corrective action whenever the test requirement is not
a. Performance Check
(1) Adjust controls of power supply (A6) for 25-volt output.
(2) Adjust SET LINE TO 21V control for red line indication on TI meter. Digital
voltmeter (A2) will indicate between 20.5 and 21.5 volts dc.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect THERMISTOR MOUNT connector of power meter (A5) to OUTPUT
NAV connector of TI, using thermistor mount (part of A5), adapter (B3), and cable supplied
with power meter.
(2) Position TI controls as indicated in (a) through (d) below:
(a) METER switch to CAR.
(b) METER ZERO control for zero indication on TI meter.
(c) CARRIER SET control for red line indication on TI meter.
(d) MICROVOLTS control to 200 K.
(3) Turn MEGACYCLES control throughout its range while maintaining red line
indication with CARRIER SET control. Power meter will indicate between 392 and 1350
microwatts. If not, perform b(1) and (2) below.
(4) Turn MICROVOLTS control to 100K.
(5) Repeat technique of (3) above. Power meter will indicate between 113 and 313
microwatts. If not, perform b(1) through(4) below.
(6) Turn MICROVOLTS control to 10K.
5