TB 9-6625-102-24
Calibrator
TI
(mV dc)
Min (C)
Max (C)
01
-3.0
+3.0
4.096
96.70
103.3
Apply stimulus and allow 5 minutes for temperature stabilization before taking
1
measurement.
(4) TI will indicate within the limits of table 9; if not, perform b below.
(5) Repeat technique of (3) and (4) above for remaining calibrator voltage listed in
(6) Set calibrator output to minimum and disconnect equipment set-up.
b. Adjustments. Perform entire calibration procedure before performing alignment
a. Performance Check
Ω
(1) Connect Resistance Standard No. 1 CURRENT HI and LO to TI
TEMPVmV and COM inputs.
(2) Set the TI as listed in (a) and (b) below:
(a) Set rotary switch to Ω.
(b) Press the RANGE key to the 500 Ω range.
(3) Set Resistance Standard to zero and press TI NULL key.
(5) TI will indicate within the limits of table 10; if not, perform b below.
(6) Repeat technique of (2)(b) through (5) above for remaining TI ranges table 10.
Table 10. Resistance Low.
Resistance
TI
Range (Ω)
Min (Ω)
Max (Ω)
Standard
500
500
499.65
500.35
5k
5k
4.9970 k
5.0030 k
50 k1
50 k
49.970 k
50.030 k
500 k
500 k
499.70 k
500.30 k
Skip line (3) above for remaining ranges.
1
(7) Disconnect Resistance Standard No. 1 from TI and connect Resistance Standard
ΩTEMPVmV input.
No. 2 terminals 1 to TI COM input and terminal 5 to TI
(8) Press the RANGE key for a 5 MΩ range.
(9) TI will indicate within the limits of table 11; if not, perform b below.
(10) Repeat technique of (7) through (9) above for remaining TI ranges and resistance