TB 9-6625-111-24
Calibrator output
Test instrument indications
Amplitude
Range
Min
Max
0.0
V
0 Hz
4
V
0.0
0.002
3.5
V
100 Hz
3.410
3.590
3.5
V
1 kHz
3.410
3.590
35
V
1 kHz
40
V
34.10
35.90
350
V
1 kHz
300.0
V
341.0
359.0
500
V
1 kHz
300
V
485.0
515.0
(5) Set calibrator output to minimum.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator to TI A and COM inputs.
~.
(2) Set rotary switch to A
(3) Set calibrator output for 3.5 A. If TI does not indicate within limits specified in
(4) Repeat technique of (3) above, using remaining settings and indications listed in
table 5. If TI does not indicate within limits specified in table 5, perform b below.
Calibrator Output
Test instrument
(Dc)
Range
Min
Max
3.5 A
4A
3.460
3.540
10.0 A
10 A
9.88
10.12
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator to TI A and COM inputs.
(2) To access A~, press and hold rotary switch button for 2 seconds.
(3) Set calibrator output for 10 A at 1 kHz. TI will indicate between 9.73 and
10.72 A ac. If TI does not indicate within limits specified, perform b below.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set TI rotary switch to Ω.
and COM inputs.
(2) Short TI V Ω
(3) TI will indicate between 000.0 and 000.2 Ω. Add this reading where indicated in