TB 9-6625-113-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Decade Capacitor, IET, Model HACS-Z-B-6E-1pF and HACS-Z-B-3E-1F. The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. Variations among models are described in text, tables, and figures.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Capacitance
Model: HACS-Z-B-6E-1pF
Range: 1 pF to 1.111,10 F
Frequency: 1 kHz
Accuracy: (0.1%+ 1.0 pF)
Model: HACS-Z-B-3E-1F
Range: 1 F Step
Accuracy: 0.1%
Frequency: 1 kHz
Range: 10 F Step
Accuracy: 0.1%
Range: 100 F Step
Accuracy: 0.5%
Range: 1000 F Step
Accuracy: 10%