TB 9-6625-116-24
(c) [
/ 1.0] (select 1.0)
(d) [CHANNEL SELECT]
(e) [LOAD 1 M ]
(f) [SIGNAL CH 1]
(g) [EXIT]
(h) [WAVE FORM]
(i) [
] (positive DC)
(j) Keypad Entry, 0.001
(k) [V]
(l) [+/-] (as necessary to select polarity of TI output in table 3)
(m) OUTPUT, ON
(4) Multimeter will indicate within the limits of table 3. If not, perform b below.
(5) Repeat technique of (3) (j) through (l) and (4) above for remaining TI outputs in
NOTE
The default high voltage interlock threshold setting is 100V.
When output exceeds 100 V, OUTPUT ON must be pressed
again to activate output. High voltage signal will sound.
TI
Output (V dc)
Min
Measured
Max
0.001
0.97475
mV
1.02525
mV
-0.001
-1.02525
mV
-0.97475
mV
0.0019
1.87453
mV
1.92548
mV
-0.0019
-1.92548
mV
-1.87453
mV
0.0023
2.27443
mV
2.32558
mV
-0.0023
-2.32558
mV
-2.27443
mV
0.005
4.97375
mV
5.02625
mV
-0.005
-5.02625
mV
-4.97375
mV
0.006
5.97350
mV
6.02650
mV
-0.006
-6.02650
mV
-5.97350
mV
0.019
18.97025
mV
19.02975
mV
-0.019
-19.02975
mV
-18.97025
mV
0.023
22.96925
mV
23.03075
mV
-0.023
-23.03075
mV
-22.96925
mV
0.05
49.96250
mV
50.03750
mV
-0.05
-50.03750
mV
-49.96250
mV
5