Adjustment - Continued
Figure 5. 100% amplitude modulation with correct offset.

Calibration Procedure For Programmable Pulse/Function Generator, Model 8116A Page Navigation
  16    17    18    19    20  21  22    23  
img
. .
TB 9-6625-127-24
Figure 4. Amplitude modulation.
ef. Disconnect TI OUTPUT from distortion analyzer and connect it to oscilloscope
input 1. Connect function generator output to oscilloscope input 2.
eg. Set TI FRQ to 15 kHz and AMP to 16 V. Change function generator amplitude to
5 V.
eh. Adjust function generator amplitude and offset until TI has 100% modulation and
minimum offset as shown in figure 5 below.