TB 9-6625-1314-35
Figure 2. Dc current - equipment setup.
(6) Repeat technique of (3) through (5) above using values fisted in table 8. If ac/dc
voltmeter does not indicate within limits specified, perform b below.
Table 8. Dc Current (Level A)
Current shunt
Test Instrument
range plugs
indication (mV)
AMPERES
RANGE
position
Switch setting
Indication
Min
Max
.001
999.9
99.78
100.21
1000 A
.01
10 MA
9.999
99.78
100.21
.1
100 MA
99.99
99.78
100.21
1
1000 MA
999.9
99.78
100.21
15