TB 9-6625-133-40
Test instrument
Performance specifications1
parameters
Accuracy
Range
(ppm of reading + ppm of range)
10 Hz - 40 Hz
145 + 25
AC Voltage -
Continued
40 Hz - 10 kHz
140 + 25
1000 V 2
10 kHz - 30 kHz
265 + 50
30 kHz - 100 kHz
700 + 250
Accuracy
Range
Mode
(ppm of reading +ppm of range)
Normal 3
22 + 2.5
2Ω
LoI 3
22 + 2.5
3
Normal
12 + 0.9
20 Ω
Lo I3
12 + 0.9
3
Normal
10 + 0.3
200 Ω
3
LoI
10 + 0.9
Normal 3
10 + 0.3
2 kΩ
3
LoI
10 + 0.9
3
Normal
10 + 0.3
20 kΩ
LoI 3
10 + 0.9
Normal
10 + 0.3
200 kΩ
LoI
10 + 0.6
Normal
12 + 0.6
2 MΩ
LoI
17 + 0.6
Normal
25 + 0.6
20 MΩ
LoI
110 + 6
Hi Voltage
20 + 0.6
Normal
150 + 60
200 MΩ
LoI
1810 + 600
Hi Voltage
80 + 60
Normal
1810 + 600
2 GΩ
LoI
1810 + 600
Hi Voltage
230 + 60
Accuracy:
Range
(ppm of reading + ppm of range)
200 uA
16 + 2.0
2mA
16 + 2.0
DC Current
20 mA
18 + 2.0
200 mA
60 + 4.0
2A
225 + 8.0
20 A
500 + 20
Accuracy
Range
(ppm of reading + ppm of range)
200 uA
10 Hz 10 kHz
620 + 120
2 mA
10 Hz 10 kHz
370 + 120
20 mA
10 Hz 10 kHz
370 + 120
AC Current
200 mA
10 Hz 10 kHz
360 + 120
10 Hz 2 kHz
725 + 120
2A
2 kHz 10 kHz
860 + 120
10 Hz 2 kHz
920 + 120
20 A
2 kHz 10 kHz
2500 + 120
See footnotes at end of table.
3